Maintenance

5
Maintenance
WARNING!
Danger from hazardous substances!
Possible consequence: Fatal or very
serious injuries.
Please ensure when handling haz‐
ardous substances that you have read
the latest safety data sheets provided
by the manufacture of the hazardous
substance. The actions required are
described in the safety data sheet.
Check the safety data sheet regularly
and replace, if necessary, as the
hazard potential of a substance can be
re-evaluated at any time based on new
findings.
The system operator is responsible for
ensuring that these safety data sheets
are available and that they are kept up
to date, as well as for producing an
associated hazard assessment for the
workstations affected.
NOTICE!
Solvent
Only use water rather than a volatile
organic solvent to clean the sensor.
1.
Clean the sensor with water and a soft
brush.
2.
Loosen tough films or deposits with
diluted hydrochloric acid and remove
with a soft brush.
26
3.
Check the complete measuring chain
(sensor, connecting line and transmitter/
controller) should there be deviations of
the measured values with an unknown
cause.