Assembly and installation

User qualification, mechanical installation:
n
trained and qualified personnel
2.2 'User qualification' on page 15
User qualification, electrical installation:
n
electrical technician
qualification' on page 15
WARNING!
Danger from hazardous substances!
Possible consequence: Fatal or very
serious injuries.
Please ensure when handling haz‐
ardous substances that you have read
the latest safety data sheets provided
by the manufacture of the hazardous
substance. The actions required are
described in the safety data sheet.
Check the safety data sheet regularly
and replace, if necessary, as the
hazard potential of a substance can be
re-evaluated at any time based on new
findings.
The system operator is responsible for
ensuring that these safety data sheets
are available and that they are kept up
to date, as well as for producing an
associated hazard assessment for the
workstations affected.
Installation site
Ensure that there is unimpeded access for sub‐
sequent calibration.
The mounting must be secure and low-vibra‐
tion.
Avoid direct sunlight.
Ä Chapter
Ä Chapter 2.2 'User
Assembly and installation
Make sure that the medium flows correctly
through and around the sensor.
Maintain a minimum distance of 20 mm from
the sensor to the tube wall when installing in
pipework.
Minimum compensation will be achieved with
the "Installation factor" parameter if you cannot
adhere to these minimum spacings.
Provide a representative installation site for typ‐
ical conductivity and/or concentration when
immersing in a tank.
19