37
Guidance and Manufacturer's Declaration – Electromagnetic Immunity for Equipment
and Systems (from RF wireless communication equipment)
For the immunity to proximity fields from RF wireless communications equipment, the DEVICE
is compliant with the test levels specified below, according to IEC60601-1-2 Table 9. The
customer or user of the DEVICE should assure that it is used in such an environment.
Test Frequency (MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
Accessories
The use of accessories other than those specified, except for those sold by the manufacturer
of the equipment, as replacement parts for internal components may result in increased
emissions or decreased immunity of the medical equipment.
Other Equipment
WARNING: Use of this equipment adjacent to or stacked with other equipment
Band (MHz)
380 – 390
430 – 470
704 – 787
800 – 960
1700 – 1990
2400 – 2570
5100 – 5800
Regulatory Information
Immunity Test Levels
Pulse modulation 18Hz, 27V/m
FM, ±5 kHz deviation, 1 kHz sine,
28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m