Neoss NeoScan 200 Gebrauchsanweisung Seite 28

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Safety Guide
WARNING: Do not rinse.
Allow to air dry.
After the scanner has dried, use a clean, lint-free cloth dampened with water to remove
residual disinfectant from the surface of the scanner.
Clean and Sterilize the Scanner Tip
WARNING
Wear gloves when handling a contaminated Scanner Tip.
Read and follow the warnings and personal protection instructions provided in the
manufacturer's SDS for the detergent used to clean the Scanner Tip.
Do not soak the Scanner Tip in disinfectant for a long period.
Dry the Scanner Tip thoroughly before mounting onto the scanner.
Do not use an ultrasonic cleaning machine to clean the Scanner Tip.
Do not soak the Scanner Tip with alcohol-based disinfectants.
Clean the Scanner Tip
To manually cleaning the Scanner Tip, follow these steps:
Rinse excess soil from the Scanner Tip (2 minutes).
Using a soft brush, apply enzymatic detergent solution (e.g., Metrex EmPower) to all
surfaces.
Rinse under clean, running water (2 minutes).
Inspect Scanner Tip. If the Scanner Tip is not clean, repeat the last three steps.
Use a lens tissue or lint-free cloth to remove any dust from the mirror in the Scanner Tip.
Sterilize the Scanner Tip
Scanner Tips shipped non-sterile. You must sterilize them before use.
CAUTION: If you limit the exposure time at 134° C to not more than 6 minutes, you can
autoclave the Scanner Tip up to 60 cycles.
To sterilize the Scanner Tip, follow these steps:
Put the Scanner Tip into a sealed steam sterilization pouch.
Place the Scanner Tip in a steam autoclave for sterilization.
For TP202 and TP203, exposure temperature should be set as 134° C.
28