Herunterladen Diese Seite drucken

Neoss NeoScan 200 Gebrauchsanweisung Seite 27

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

27
You must wear gloves while cleaning and disinfecting the scanner.
The scanner must be disinfected with a recommended intermediate-level disinfectant
solution with tuberculocidal activity between patients.
DO NOT use a disinfectant containing phenolics or iodophors; doing so will damage the
surface coating of the scanner.
Never put the scanner in an autoclave device or immerse it in water or the disinfectant
solution.
Excessive fluids can damage the scanner.
Do not use cotton, cloth, or tissues soaked with disinfectant to disinfect the scanner.
Clean the Scanner
If the scanner is visibly contaminated with blood and/or body fluids, you must clean it
before disinfecting it.
To clean the scanner, follow these steps:
Dampen (do not soak) a lint-free cloth with lukewarm water.
Remove the blood and/or body fluids with the dampened lint-free cloth.
Disinfect the Scanner
After each patient, the scanner must be thoroughly disinfected.
To adequately disinfect the scanner, follow the disinfectant manufacturer's instructions for
the appropriate contact time.
CAUTION: If the scanner is visibly soiled, it must be thoroughly cleaned prior to disinfecting.
See "Clean the Scanner"
To disinfect the scanner, follow these steps:
Remove the Scanner Tip.
Remove all visible soil (see "Clean the Scanner").
Use a commercially prepared intermediate level disinfectant wipe. Follow the
manufacturer's instructions for contact time.
Recommended disinfectant wipes: CaviWipes
WARNING: Using a disinfectant that has not been approved may cause damage to
the scanner.
Thoroughly wipe all surfaces of the scanner, DO NOT allow liquid to enter through the gap,
air outlet, or pin holes.
Safety Guide

Werbung

loading

Verwandte Produkte für Neoss NeoScan 200