Safety Guide
DO NOT expose the scanner to high vibrations.
DO NOT expose the scanner to ultraviolet radiation directly. The scanner is not designed
for ultraviolet disinfection.
DO NOT stare at the LED emission window.
When the Scanner Tip is removed, install the front protective cover to protect the Lens
Window.
DO NOT remove the cover of any scanner components. The scanner contains no user-
serviceable parts. For any repairs, contact a qualified Alliedstar service technician.
DO NOT replace the cables provided with the scanner with other cables. Doing so may
damage the scanner and adversely affect the safety protection and EMC performance of
the scanner.
Any other equipment not complying with IEC-60601 shall be kept at least 1.5 meters away
from the patient.
If the equipment is faulty, turn it OFF, display an "Out of Service" notice, and contact a
qualified Alliedstar service technician.
Using components, accessories, cables, and spare parts other than those specified
or provided by the manufacturer of this equipment may impair the safety protection
of the scanner and may result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
No modification of this equipment is allowed.
Additional multiple outlet strips or extension cords should not be connected to the system.
The maximum temperature of the applied part may reach to 43 ° C; to avoid overheating,
do not use it for extended periods.
To isolate the holder from power supply, unplug the USB connector from the USB port.
DO NOT maintain or service this equipment while it is in use with the patient.
Connection of the PEMS (Programmable Electrical Medical System) to an IT NETWORK that
includes other equipment could result in risks to patients, operators, or third parties. The
responsible organization should identify, analyze, evaluate, and control these risks.
Patients with oral mucosal disease, mental illness, severe respiratory disease, asthma,
Parkinson's disease, hyperactivity disease are forbidden.
Patients with moderate or severe opening limitation should use it with caution.
Computer
Do NOT place any equipment which does not comply with IEC 60601-1 in the immediate
vicinity of the patient. Leave at least 1.5 meters distance between the patient and the
equipment.
The scanner is only intended to be connected to a computer that is at least IEC 60950 / IEC
62368, or equivalent standards certified. Connecting the scanner to other equipment may
24