Regulatory Information
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
NOTE: Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast,
and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the DEVICE is
used exceeds the applicable RF compliance level above, the DEVICE should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the DEVICE.
IEC 60601 Test
Level
±8 kV contact
±15 kV air
30 A/m
3 V/m
80MHz – 2.7GHz
Compliance Level
±8 kV contact
±15 kV air
30 A/m
3 V/m
80MHz – 2.7GHz
36
Electromagnetic
Environment -
Guidance
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Environment
of a
professional
healthcare facility.