The C-MAC
ZX/8403 ZX as well as in conjunction with the C-MAC
customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the unit as recommended below, according to the maximum output power of the communications equipment.
Rated maximum out-
put power of trans-
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
36
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the
KARL STORZ C-MAC
®
video laryngoscope 8401 AX/BX/HX/AXC/BXC/DXC/GXC/KXC/HXP in conjunction with the electronics module 8401 X/8402 X and the monitor 8401 ZX/8402
mitter
150 kHz to 80 MHz
[W]
d=[3,5/3] P
0.01
0.1
0,1
0.4
1
1.2
10
3.7
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 6
®
video laryngoscope
8401 AX/BX/HX/AXC/BXC/DXC/GXC/KXC/HXP
®
PM 8401XD is intended for use in an electromagnetic environment in which the HF disturbances are controlled. The
Separation distance d [m]
according to frequency of transmitter
80 MHz to 800 MHz
d=[3,5/3] P
0.1
0.4
1.2
3.7
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
800 MHz to 2.5 GHz
d=[7/3] P
0.2
0.7
2.3
7.4

Werbung