Herunterladen Diese Seite drucken

HBM QuantumX MX840B Bedienungsanleitung Seite 22

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 30
Connecting individual QuantumX modules
6
Connecting individual QuantumX modules
6.1
Connecting the supply voltage
Connect the module to a DC voltage of 10 to 30 V (e.g. 24 V). If a supply
voltage of > 30 V is used, defects in the module cannot be ruled out. If the
supply voltage falls below 10 V, the module switches itself off. Reverse polarity
protection is guaranteed.
For battery operation in a vehicle, we recommend installing an uninterruptible
power supply (such as the HBM 1-UPX001-2) between the battery and the
module, to compensate for voltage dips during startup. For continuous
operation without wearing out the installed battery, we recommend a backup
battery that can be used when required.
6.2
Connection to a PC
To connect a module to a PC, use a standard Ethernet cable (e.g. HBM
1-KAB239-2) and connect it to the module and your PC's Ethernet port as
illustrated below.
The CX27C, MX471C and MXFS modules can also be connected to the PC or
fieldbus via the RJ45 ports on the front and the gateway function.
20
Or
1-Kab271-3
A02361_05_Z00_01 HBM: public
QuantumX

Werbung