2
Electro magnetic conformity
Additional information about the relevant EMC standards EN 61326-1 /
EN61326-2-x.
These standards define emissions limits and immunity requirements for differ­
ent environments.
Emissions requirements are defined for the following environments:
- Industrial (Class A) or
- Residential / Laboratory (Class B).
The standard refers to CISPR 11:2009+A1:2010.
Immunity requirements are defined for the following environments:
- Controlled electro-magnetic (lowest requirements)
- Basic or
- Industrial (highest requirements).
The modules listed in the declaration of conformity comply with the require­
ments for the following environments:
Emissions: Class A
Immunity: Industrial environment
The QuantumX series and its modules are intended for use in an industrial
environment. When used in residential or commercial environments, additional
arrangements may be required to limit electro-magnetic emissions.
An example is voltage supply of the modules by battery. In this case please
wrap the power supply cable (KAB271-3) around the inductive coil included in
the package four times.