Herunterladen Inhalt Inhalt Diese Seite drucken

Changing Between Celsius And Fahrenheit; How To Recall 12 Readings In Memory Mode; Error Messages - Microlife IR 120 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 68
3. When the «°C» or «°F» icon is flashing, a beep sound is
heard and the thermometer is ready for measuring 8.
4. Straighten the ear canal by pulling the ear up and back
to give a clear view of the eardrum.
 For children under 1 year; pull the ear straight back.
 For children of 1 year to adult; pull the ear up and back.
Also refer to the short instructions at the front.
5. While gently pulling the ear, insert the probe snugly into
the ear canal (about 1 sec.) and immediately press the
START button 2. Release the button and wait for the
beep sound. This indicates the end of the measurement.
6. Remove the thermometer from the ear canal. The display
shows the measured temperature 9.
NOTE:
 To ensure accurate readings, wait at least 30 sec.
after 3-5 continuous measurements.
 Accumulation of ear wax on the probe can result in less
accurate temperature readings or cross infection between
users. Therefore, it is essential that the probe is clean
before each measurement. For cleaning, follow the
instructions in the «Cleaning and Disinfecting» section.
 After cleaning the measuring sensor 1 with alcohol,
wait 5 minutes before taking the next measurement, in
order to allow the thermometer to reach its operating refer-
ence temperature.
IR 120
 10 short beeps and a red LCD backlight alert the patient
that he/she may have a temperature equal to or higher
than 37.5 °C.
 For an infant, it is best to have the child lying flat with his/
her head sideways so the ear is facing upwards. For an
older child or adult, it is best to stand behind and slightly
to the side of the patient.
 Always take the temperature in the same ear, since the
temperature readings may be different from ear to ear.
 In the following situations it is recommended that three
temperatures in the same ear be taken and the highest
one taken as the reading:
1. New born infants in the first 100 days.
2. Children under three years of age with a compromised
immune system and for whom the presence or absence
of fever is critical.
3. When the user is learning how to use the thermometer
for the first time until he/she has familiarized himself/
herself with the device and obtains consistent readings.
4. If the measurement is surprisingly low.

6. Changing between Celsius and Fahrenheit

This thermometer can display temperature readings in either
Fahrenheit or Celsius. To switch the display between °C and
°F, simply turn OFF the unit, press and hold the START
button 2 for 5 seconds; after 5 seconds, the current
measurement scale («°C» or «°F» icon) will flash on the
display AL. Change the measurement scale between °C and
°F by pressing the START button 2. When the measure-
ment scale has been chosen, wait for 5 seconds and the unit
will automatically enter the «ready for measuring» mode.

7. How to recall 12 readings in Memory Mode

This thermometer can recall the last 12 readings.
 Recall mode AM: Press the START button 2 to enter
Recall mode when power is off. The memory icon «M»
will flash.
 Reading 1 - the last reading AN: Press and release the
START button 2 to recall the last reading. Display 1
alone with memory icon.
 Reading 12 - readings in succession: Press and
release the START button 2 consecutively to recall the
last 12 readings in succession.
Pressing and releasing the START button 2 after the last
12 readings have been recalled will resume the above
sequence from reading 1.

8. Error Messages

 Measured temperature too high AO: Displays «H» when
measured temperature is higher than 100.0 °C / 212.0 °F.
5
EN

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis