Herunterladen Inhalt Inhalt Diese Seite drucken

Connection J1 - Logic Input Interconnections - Agilent Technologies Turbo-V 2000 ICE Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
17
Technical Information
Interconnections

Connection J1 - Logic Input Interconnections

Figure 8
All the logic input to the controller must be connected at J1 mating
connector. With the provided J1 mating connector (shipped with pin
3 and pin 8 shorted) make the connections with AWG 24, (0.24 mm2)
or smaller wire to the pins indicated in the figure to obtain the
desired capability. The following table describes the signals available
on the connector.
212/260
FEMALE CONNECTOR
13
25
12
24
11
23
10
22
9
21
8
20
J1
7
19
I/O
6
18
5
17
4
16
3
15
2
14
1
OPEN COLLECTOR OUTPUTS IMAX = 60mA
POWER MONITOR : 10V = 1000W
SPEED MONITOR : 10V = 33KRPM
J1 input connector
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
SPEED MONITOR (0 TO 10V)
GND
POWER MONITOR (0 TO 10V)
BRAKING (OPEN COLLECTOR)
FAILURE (OPEN COLLECTOR)
+24VDC
OVERTEMP (OPEN COLLECTOR)
+24VDC
PURGE FLOW (OPEN COLLECTOR)
+24VDC
NORMAL S2 (OPEN COLLECTOR)
+24VDC
NORMAL S1 (OPEN COLLECTOR)
+24VDC
STARTING S2 (OPEN COLLECTOR)
+24VDC
STARTING S1 (OPEN COLLECTOR)
+24VDC
REMOTE/LOCAL (OPEN COLLECTOR)
GND
INTERLOCK
GND
START S2
GND
START S1