Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies Turbo-V 2000 ICE Bedienungshandbuch Seite 239

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
Select the type of purge gas, the default is Nitrogen. Argon can be
seleted if needed.
S
Heater Jacket power ON or OFF: setting ON powers on the Heater
Jacket, setting OFF powers it off. The Heater Jacket automatically
switches to OFF when the controller is in the Failure State.
H
S
Heater jacket operating temperature selection. The range is 40-90 °C,
set to 65 °C by default.
T
Detection of the pump's lower bearing temperature. OFF will always
be indicated if there is no sensor on this bearing.
and then:
S
S
Where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to rump-up
the Normal speed within ten steps.
When NO is selected, the Soft Start mode is deselected and the rump-
up of the pump will be done within 10 minutes.
The controller is factory set to YES.
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
Press the PUMP CURRENT push-button and the display shows:
H
E
A
T
E
E
L
E
C
T
Press the PUMP CURRENT push-button and the display shows:
E
A
T
J
E
T
T
I
N
Press the PUMP CURRENT push-button and the display shows:
L
O
W
E
R
2
S
E
N
O
F
T
S
E
L
E
C
T
Technical Information
R
J
A
C
I
O
N
:
A
C
K
E
T
G
:
X
X
B
E
A
R
S
O
R
:
T
A
R
T
I
O
N
:
17
Use
K
E
T
O
N
T
E
M
P
°
C
I
N
G
O
F
F
M
O
D
E
X
X
X
239/260

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

969-9448969-9449

Inhaltsverzeichnis