Herunterladen Diese Seite drucken

Input/Output Interconnections; Input Signals - Agilent Technologies MicroVac 929-0200 Bedienungshandbuch

Werbung

5
Technical Information
Fuse Holder and Voltage Changer Assembly
Figure 12
6.

Input/Output Interconnections

All input/output signals are accessible at J001 accessory connector.
With the provided P001 mating connector make the connections with
AWG 24 (0.24 mm
following figure, to obtain the desired capability.

Input Signals

Pin 3-7 Remote high voltage ON/OFF or high voltage interlock;
Pin 3-7 must be shorted to allow the high voltage output if no remote interlock
NOTE
contact is connected.
58/72
Voltage Changer
Check voltage selector window for correct set and connect the
power cord.
2
) - or smaller wire - to the pins indicated in
requires a permanent closed pin contact.
MicroVac Controller User Manual / 87-900-089-01 (G)

Werbung

loading

Diese Anleitung auch für:

Microvac 929-0201