Herunterladen Inhalt Inhalt Diese Seite drucken

Window-Based Protocol - Agilent Technologies Turbo-V 2000 ICE Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
17
Technical Information
Examples

Window-Based Protocol

The following table, valid for the RS 232, RS 422 and RS 485
interfaces, describes each single window used in the protocol.
Tab. 7
WIN
000
001
100
102
107
108
109
110
111
112
113
114
115
116
117
118
119
200
201
202
203
205
230/260
R
W
T
Description
X
X
L
START/STOP
X
X
L
SPEED SELECTION ACTIVE [0=HS / 1=LS]
X
X
L
SOFT START [0=NO / 1=YES]
X
X
L
WATER COOLING [0=NO / 1=YES]
X
X
N
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
X
X
N
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
X
L
PUMP LIFE RESET AND FLOW METER ALARMS (TYPE
"ON" TO RESET)
X
X
N
THRESHOLD HS [Krpm]
X
X
N
THRESHOLD LS [Krpm]
X
X
L
HEATER JACKET ENABLE [0=ON / 1=OFF]
X
X
N
HEATER JACKET TEMPERATURE SETTING [Degrees]
X
X
L
GAS TYPE [0=N
X
X
N
THRESHOLD FLUSSIMETRO [sccm]
X
X
N
HIGH SPEED ADJUST [Krpm]
X
X
N
LOW SPEED ADJUST [Krpm]
X
X
L
LOWER BEARING TEMPERATURE MEASUREMENT
ENABLE [0=NO / 1=YES]
X
X
L
HOST / PRINTER MODE [0=HOST / 1=PRINTER]
X
N
CURRENT [A]
X
N
VOLTAGE [V]
X
N
POWER [W]
X
N
FREQUENCY [Krpm]
X
N
PUMP STATE (0÷6) [STOP, WAITING INTERLOCK,
STARTING, NORMAL, HIGH LOAD, FAILURE,
APPROACHING]
Turbo-V 2000 ICE Controller User Manual / 87-900-929-01 (L)
/ 1=Ar]
2

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

969-9448969-9449

Inhaltsverzeichnis