3. Do not under fill the disk. If the disk is under filled,
the blank chamber will not be full. The reagent
chambers will not fill entirely and results will be
inaccurate.
4. Do not introduce air bubbles into the disk. The
reagent chambers will not fill entirely and results
will be inaccurate. As soon as a bubble starts to
form, pull back on the plunger to draw the bubble
out of the disk. Begin the filling process again.
5. Wet disks should be dried thoroughly with a lint
free wipe. Handle the disk by the edges. Between
tests store the disk on a flat surface. Do not
agitate a partially filled disk to avoid leaking and
contamination. Partially used disks should be
discarded at the end of the day.
6. Disks should be filled and used within 10 minutes.
They cannot be filled ahead of time.
For filling and troubleshooting tips go to:
GENERAL OPERATING PROCEDURES
METER
When a filled disk is placed in the chamber and the lid is closed, the meter
spins at high speed to distribute the sample to the test wells. Next the meter slows
to maximize the pumping action of the stainless steel mixing beads as the reagents
mix with the sample water. Each reaction is then read at the proper time and
wavelength for that reagent system.
The button located in the lower center of the top of the meter turns the
instrument on and off.
The Blue indicator light of the on/off button indicates the status of the instrument.
Steady blue light –
the blue LED will remain steady to indicate that the meter is on
and ready to run a test.
Blinking blue light
(three blinks/second) – a test is in progress and the disk is
spinning. Do not open the lid when the disk is spinning.
Care should be taken when closing the lid. Do not slam the lid. Wiring between the
lid and the body of the photometer passes through the hinge. The meter will not run
with the lid open.
TOUCHSCREEN DISPLAY
When the button is pressed to turn the meter on, the Test Screen will be displayed.
The display screen is touch-activated. To make a selection, tap the icon or word on
the screen with a fingertip, fingernail, pencil eraser, or stylus.
¬ Gently wipe smudges from the screen with the Cloth Wipe (3580-WIPE).
¬ Do not touch the screen with a sharp object.
¬ Do not place objects on the screen that will scratch or damage it.
¬ Avoid touching the screen with wet fingers.
TESTING
1. Press
and hold until
the meter turns on.
2. Tap
. Select a sanitizer
type. Tap
to confirm.
3. Tap
. Select a disk
series (found on disk
packaging). Tap
to
confirm.
Disk
Type
NOTE: Disk Series are
limited by Sanitizer
selection.
Power/
4. Remove a disk from the
Battery
packaging.
Status
5. Use the syringe (1189) to
fill the disk with the pool
or spa water sample.
6. Insert the disk. Cover the
disk with the Universal
Disk Cover (1719). Close the lid.
7. Tap
to start test. Tap
to cancel the test. If the test is cancelled discard
the disk.
8. The results will be displayed.
9. Choose an option.
¬ Tap
to add a tag.
¬ Tap the highlighted
to save the test results to the test log if Auto Save is
not enabled.
¬ Tap the hightlighted
to send the results to the enabled Mobile Bluetooth
Printer.
¬ Tap the highlighted
to send the results to a Bluetooth enabled device.
¬ Tap
to return to the Test Screen.
Blank
Chamber
Underfilled
Air Bubble
lamotte.com/spin_support
Start
Sanitizer
Test
Sanitizer
Salt
Tap to
Start Test
Disk Series
203
16-Apr-20
12:14 PM
Date &
Test
Time
History
10. REMOVE THE DISK FROM THE CHAMBER. DISPOSE OF THE DISK. Disks left in the
chamber can leak and damage the device. Do not store the disk cover on the
hub.
11. Press and hold
NOTE: For water samples over 100 °F (38 °C) subtract 0.1 from pH result or, for the
most accurate result, wait until water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results.
See Cleaning.
SAVE, PRINT, AND SEND TEST RESULTS
Test results can be saved, transferred to
the WaterLink Connect 2 mobile app via
Bluetooth, and sent to the BLE Mobile Printer
(Code 5-0067).
Test History Settings
Touch can log test results for 250 water
samples in the Test History. The results for
the most recent sample will be located at
the top of the list. All results can be logged
automatically or results for an individual
sample can be logged after the sample has
been tested.
To turn on automatic logging, tap
Test Screen. Tap
. Select Auto-Save Tests.
Tap
and
to return to the test screen. When Auto Save is selected
highlighted on the Test Results screen.
To manually log results for one sample at a time, Auto-Save Tests must be disabled.
If Auto-Save Tests is disabled
the test has been run, tap
History.
A user defined identification tag can be assigned to an individual sample result.
To enable tagging, tap
and
to return to the test screen. The tag can be up to 4 hexadecimal characters.
It cannot begin with a zero. To tag a test result, tap
Results screen and use the buttons to enter up to four characters. Tap
a character. Tap
to save the tag and return to the Test Results screen. Tap
to return to the Test Results screen without saving the tag. The tagged result
will appear on the Test Results screen and the Test History screen. The tag will be
preceded by "T-".
Logged results are viewed in Test History. Controls for viewing and managing single
or multiple test records are located on the Test History screen. Tap the checkbox
next to a test record to select it, then tap one of the buttons along the bottom to
perform an action with the selected records.
Print or Send Results
to a Bluetooth Phone
or Tablet
The WaterLink
Spin Touch can print to
the Mobile Bluetooth
Printer or send test
results to a Bluetooth-
enabled device such as
a phone or a tablet. The
highlighted
or
will show which function
is active. The Spin Touch
cannot connect to the
printer and connect to
a phone or tablet at the
same time.
There are two modes of operation for connecting to the printer and a phone or tablet
– Fast Printer Connect ON and Fast Printer Connect OFF. The default mode is Fast
Printer Connect ON. Fast Printer Connect can be turned on or off from the Bluetooth
Menu that is located in the Settings Menu.
If Fast Printer Connect is ON, the meter will connect immediately to the printer upon
recognizing it. It will also connect to a phone or tablet but priority is given to the
printer connection. The printer will start printing immediately when
The
will dim in intensity while the printer is printing. Fast Printer Connect should
be ON if results will be printed more frequently than results will be sent to a phone or
tablet. It can remain selected if results will only be sent to a phone or tablet. If Fast
Print Connect is ON, and the printer is on, the printer will have to be turned off before
results can be sent to a phone or tablet.
If Fast Printer Connect is OFF, the meter has the opportunity to connect to a phone
Settings
or tablet at all times other than when the printer is actually printing a test. The meter
will connect to the printer only after
few seconds while the connection to the printer is being established. The
in intensity while the printer is printing. After printing, the meter will automatically
reconnect to the phone or tablet. Fast Printer Connect should be OFF if results will be
sent to a phone or tablet most frequently and printed less often.
SYRINGE
A plastic 3 mL syringe (Code 1189) is used to fill the disks. A precision
tip on the syringe fits into the fill hole on the disk. The syringe tip should not be
removed from the syringe. Syringes should be cleaned between water samples.
Pump air in and out of the syringe a few times to clear the previous sample or
rinse the syringe with a small amount of the next water sample before filling
it with the next sample. Replace syringes when the tips become worn or the
plungers don't move smoothly. See Accessories and Replacement Parts.
3
for 2 seconds to turn the meter off.

Test Results

Test Time 05:30:06 16-Apr-2020
FCL - 1.31
TCL - 1.31
CCL - 0.00
The WaterLink Spin
PH - 7.8
ALK - 100
HARD - 230
Save
Tagging
Results
on the
will be highlighted on the Test Results screen. After
to save the results for that water sample to the Test
on the Test Screen. Tap
. Select Enable tags. Tap
on the bottom of the Test
Test History
12:02A 16-Apr Salt 401
12:01A 16-Apr Salt 301
12:00A 16-Apr T-4650 Chlorine 100
12:00A 16-Apr Chlorine 300
11:57A 16-Apr T-ABC7 Chlorine 300
11:49A 16-Apr Biguanide 400
Select All/
View
Delete
History
De-Select All
Selected
Selected
Settings
Results
Results
Select
Auto Save
has been tapped so printing will be delayed a
CYA - 2
COPPER - 0.1
IRON - 0.1
PHOS - 0
SALT - 489
Print
Transfer
Return
Results
Results
to Test
via
via
Screen
Bluetooth
Bluetooth
will not be
to erase
Go to Top
Scroll up
or down
the test
list by 5
entries
Go to Bottom
Home
is tapped.
will dim