Figure 2-32: Removing the Summing Board
The summing board can then be placed on the bottom of the enclosure for wiring,
as in Figure 2-33.
Figure 2-33: IND131 Terminal Removed, Summing Board in Position for Wiring
Prepare the ends of each of the load cell cables as shown in Figure 2-34 and
Table 2-4. Make sure the shield is connected to the gland as depicted in Figure
2-26.
A
Figure 2-34: Load Cell Cable Preparation
Table 2-4: Load Cell Cable Length
Cable
Shield Length (A)
Load Cell 1
Load Cell 2
Load Cell 3
Load Cell 4
IND131/IND331 Terminals Installation Manual
B
Stripped Cable Length (B)
25mm (1 in.)
25mm (1 in.)
305mm (12.0 in.)
25mm (1 in.)
25mm (1 in.)
Summing board
mounting studs
205mm (8.1 in.)
160mm (6.2 in.)
135mm (5.3 in.)