• The inlet zones can be reduced by additionally installed flow conditioners.
Flow conditioners create an additional pressure loss that can reduce the flow
quantity.
Table 3: Installation direction
Type
3
1
2
Procedure
Fig. 5: Assembly (pipe product variant)
1. Align the product observing the flow direction of the measuring medium (in-
put left, output right).
2. Connect the product to the pipe on both sides.
3. Pressurize the line section.
4. Check the installation for tightness and correct functioning.
AVENTICS™ 652/653 | 549411-001-AB | English
Pipe diameter = DN (3)
Inlet zone (1)
Expansion
10 x DN
Constriction
10 x DN
90° bracket
20 x DN
2x 90° brackets
25 x DN
on 1 level
2x 90° brackets
30 x DN
on 1 level, T-con-
nector
Valve, pusher
50 x DN
2.
2
1.
1
1.
1
2.
2

6.3.2 Assembly with mounting bracket

Outlet zone (2)
3 x DN
3 x DN
5 x DN
5 x DN
5 x DN
Fig. 6: Assembly with mounting bracket
1. Align the product observing the flow direction of the measuring medium (in-
put left, output right).
2. Mount the product with a mounting bracket. See g Fig. 6.
3. Pressurize the line section.
5 x DN
4. Check the installation for tightness and correct functioning.

6.3.3 Assembly with block assembly kit

Fig. 7: Assembly with block assembly kit
Series
652
653
1. Align the product observing the flow direction of the measuring medium (in-
put left, output right).
2. Connect the product to the block assembly kit that matches the line.
3. Mount the product with the block assembly kit. See g Fig. 7.
4. Pressurize the line section.
5. Check the installation for tightness and correct functioning.

6.4 Connecting the electrical system

The variant with IO-Link features 1 of the following options:
• 2x switch output
• 1x IO-Link communication
• 2x analog outputs
The switch outputs at contacts 5 and 4 can be used as type NO, NC and wired ac-
cordingly. See
2
Size
Torque
1
3
2.9 ± 0.2 Nm / 25 ± 2 inch-pounds
2
4
4.2 ± 0.2 Nm / 37 ± 2 inch-pounds
1
3
2.9 ± 0.2 Nm / 25 ± 2 inch-pounds
2
4
4.2 ± 0.2 Nm / 37 ± 2 inch-pounds
data.