Connection with a 6‐wire cable
Schematic diagram of a TE connector (TE 3-640442-6), 6-pin
5.2

Cable extensions

Use only shielded, low-capacitance measurement cables to extend cables, making sure
there is a proper connection with minimal contact resistance.
The cable of a six-wire load cell can be extended using a cable of the same type.
The cable of a four-wire load cell should be extended with a 6-wire cable. At the
connecting point between the 4-wire and 6-wire cables, the supply leads must be
connected to the relevant sense lead.
5.3

Shortening the cable

The cable of a load cell with a six-wire configuration can be shortened without this
changing the calibration.
The cable of a load cell with a four-wire configuration should not be shortened, as the
particular cable length is taken into consideration in the calibration.
14
Plug-in contact 4 (blue [BU]) = excitation voltage (+)
Plug-in contact 5 (green [GN]) = sense line (+)
Plug-in contact 1 (white [WH]) = measurement signal (+)
Plug-in contact 3 (black [BK]) = excitation voltage (-)
Plug-in contact 6 (gray [GY]) = sense line (-)
Plug-in contact 2 (red [RD]) = measurement signal (-)
Shield (yellow [YN]) = Cable shield