2. The blue Sternum Relief Knob adds a degree of freedom for the
patient so that activities such as eating or talking can be performed
with greater comfort. The knob turns 90 degrees to a horizontal
position. (Fig. 10)
3. Upon physician recommendation, collar corners and edges may be
modified or trimmed to relieve discomfort or pressure.
Final Fitting Checklist
A properly applied Miami J Select will look like Fig. 11.
• Collar extends from mandible to just below sternal notch.
• Chin is centered comfortably in Chin Support. Chin should not extend
over edge of Sorbatex™ pad, nor fall inside collar.
• Straps are aligned to the Hook landing area and are an equal length.
• Sides of Back should overlap sides of Front.
• Front of collar angled up toward ears. Lower plastic edge should not
be resting on patient´s clavicles nor digging into trapezius.
• No plastic touching skin. Sorbatex™ pads extend beyond all plastic
edges.
• No slack or gaps in any of the straps.
• Tracheal opening and posterior vent are midline.
• Collar sits away from the neck. There should be a "finger" gap
between the tracheal opening and the neck. If the collar is fit too
closely/tightly to the neck, size up to the next taller size.

ACCESSORIES AND REPLACEMENT PARTS

Please refer to the Össur Catalog for a list of available replacement parts
or accessories.
CLEANING AND CARE (Fig. 12)
• Remove the pads from the plastic shells.
• Hand wash the pads with mild soap and water. DO NOT use harsh
detergents or bleach.
• Rinse. Wring out the excess water and squeeze in towel. Lay flat to air
dry.
• Wipe the plastic collar shells clean with mild soap and water.
• Replace pads by attaching the grey side to the hook on the inside of
the collar.
DISPOSAL
The device and packaging must be disposed of in accordance with
respective local or national environmental regulations.
LIABILITY
Össur does not assume liability for the following:
• Device not maintained as instructed by the instructions for use.
• Device assembled with components from other manufacturers.
• Device used outside of recommended use condition, application, or