15. Use of the microscope (B-510FL/B-510LD4/B-510LD4-SA/B-510LD4D)

This section refers exclusively to the use of the reflected light fluorescence microscope.
For transmitted light operations, refer to this manual in chapters 9-10-11-12.

15.1 Microscope setting (B-510FL)

Centering the HBO mercury bulb.
Wait around 5 minutes before proceeding with this
operation to allow the bulb to warm up properly.
1. Turn on the mercury bulb by operating the power supply
main switch ①. (Fig. 85)
2. Turn the nosepiece into an empty position (without objectives)
and remove the protective cap, or remove an objective from
the nosepiece.
3. Place a piece of white paper on the stage and insert the
fluorescent cube "B" into the optical path. (Fig. 86)
4. Acting on the focus screw of the collector lens ② and on the
centering screws ③ try to obtain the light spot of the bulb's
arc. (Fig. 87-88)