8.11 Polarizing set (optional)

1. Place the polarizer on the light exit ① at the base of the mi-
croscope. (Fig. 49)
2. Loosen the head fixing knob ② and remove the head from
the microscope frame. (Fig. 50)
3. Insert the analyzer into the hole inside the frame ③. (Fig. 51)
4. Put back the head into its original position and lock the fixing
knob.
The use of the polarization set, although possible for
models B-510FL, B-510LD4, B-510LD4SA and B-510LD4D,
is not recommended. The presence of the analyzer within
the optical path, during the use of fluorescence, causes
a significant reduction in the amount of light projected
on the sample, resulting in difficulty of observation.