! WARNING: Point the beam of light directly onto the material to be polymerised. Do
not subject the gum or other soft tissues to the beam of light (shield these parts suitably if
necessary). the effect of the light should be limited to the oral cavity and in particular to the
sector requiring clinical treatment..
! WARNING: Do not short-circuit the electric contacts of the handpiece with metal objects
or liquids (Figure 6 at page 14 - Ref. A and Ref. B).
CAUTION: During the first few seconds of exposure avoid contact of the tip with the
material to be polymerised. Deposits of composite material adhering to and polymerised
to the tip terminal surface lower the amount of light transmitted and will therefore prejudice
subsequent polymerisation operations.
! WARNING: replace damaged or inefficient optical fibre as the light intensity is
significantly reduced.
! WARNING: During the intervention on the patient, do not perform any maintenance tasks
on the system.
6.3
Instructions for Use
starlight pro allows performing 2 types of exposures:
• FAST: exposure time of 10 seconds at the maximum light intensity.
• SLOW RISE: exposure time of 20 seconds with a gradual increase of the light intensity
during the first 3 seconds up to the maximum intensity.
FAST exposure selection:
1.
Press the on/off button on the handpiece (Table 1 at page 11 - Ref. D) for a short time.
An acoustic signal will be emitted (1 beep).
2.
After 10 seconds an acoustic signal will be produced (1 beep). The FAST cycle has been
completed.
SLOW RISE exposure selection:
1.
Keep pressed the on/off button on the handpiece (Table 1 at page 11 - Ref. D) for 2
seconds. An acoustic signal is emitted when the cycle starts and after 2 seconds.
2.
After 10 seconds an acoustic signal will be emitted (1 beep).
3.
After 20 seconds an acoustic signal will be emitted (1 beep). the SLOW rISE cycle has
been completed.
After the end of the treatment, place the starlight pro handpiece back into the charging unit
(Figure 4 at page 8 - Ref. B).
NOTE: Cycle interruption.
Both in the FASt and in the SLOW rISE mode, the exposure cycle can be interrupted at any
time by pressing the on/off button on the handpiece (Table 1 at page 11 - Ref. D).
NOTE: Additional exposures.
At the end of any exposure cycle, it is possible to carry out one or more additional cycles by
pressing the on/off button on the handpiece again each time (Table 1 at page 11 - Ref. D).
For a quick guide to the signalling, see Table 2 at page 12 and Table 3 at page 13.