Herunterladen Inhalt Inhalt Diese Seite drucken

Waste Water Connection - Fagor ADVANCE AG-061 Installations, Bedienungs Und Wartungsanleitungen

Vorschau ausblenden Andere Handbücher für ADVANCE AG-061:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 91
EN
Only UK: IRN R160
To be carried out by the installer: a certified double check valve, or other device suitable for
preventing the return of the water, of at least "fluid category 3", must be fitted on each of the
drinking water connections to the appliance.
Never use detergents which represent a risk higher than the category 3 fluids.
Mount a main tap on each appliance.
Rinse the water pipe before connecting it to the appliance.
3.8

WASTE WATER CONNECTION

Before connection, look at the label that
shows which is the water inlet.
An inappropriate installation of the oven may lead to an undesirable
operation of the device.
Therefore, you must install the drain KIT (DN50) supplied by the
manufacturer (Fig. 12). This drain must preferably be connected to a
grid or an open tank.
Installation (Fig. 13- Fig.14.) should be in such a way as to ensure that
the installed drainage outlet is below the oven outlet with a suitable
slope to ensure drainage (>5% or 3º).
Ensure that the holes of the rise tube are directed towards the opposite
side of the back panel to avoid condensation.
Observe the dimensions for the drain:
Steam generator pumping volume in a reduced space of
time: 0.7 l/s.
Average temperature of waste water: 65 ºC.
78

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis