Technical data
Influence of ambient temperature (temperature drift)
• Pt100 resistance thermometer:
= ±(15 ppm/K * (full scale value of measuring range + 200) + 50 ppm/K * set measuring range) *  
T
d
  = deviation of ambient temperature from the reference operating condition.
Influence of load
± 0.02%/100 . Specifications refer to the full scale value of the measuring range.
Transmitter response time
1 s
Sensor response time (measured as per IEC 60751, in water flowing at 0.4 m/s (1.3 ft/s)
< 1.0 s
t
50
Influence of supply voltage
 ±0.01%/V deviation from 24 V. Specifications in percent refer to the full scale value of the measuring range.
Self-heating
Negligible small
Sensor current
< 0.6 mA
9.5

Mechanical construction

Design, dimensions
See Technical Information Easytemp
PDF download at: www.products.endress.com/tmr31 or www.products.endress.com/tmr35
Weight
Version with L = 100 mm (3.94 in)
TMR31 with G½", ISO 228 process connection
TMR35 with ISO2852 DN25-38, with clamp process connection (DB) 1 to 1½"
28
t
90
®
TMR31, TMR35 (TI123R/09/en),
< 2.0 s
116 g (4.1 oz)
305 g (10.76 oz)