Stage III
Fully immerse the product in the disinfectant
solution.
Rinse lumens at least 5 times at the beginning of
the exposure time, using a disposable syringe
(20 ml) and an appropriate rinsing adapter. Ensure
that all accessible surfaces are moistened.
Stage IV
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The disinfector must be of tested and approved
effectiveness (e.g. DGHM or FDA approval or CE mark
according to DIN EN ISO 15883).
Note
For thermal disinfection, always use fully desalinated
water (demineralized, low microbial contamination:
drinking water quality at least) and ensure that Ao is
>3 000 for the process.
Note
The disinfector used for processing must be serviced
and checked at regular intervals.
Rinse the lumens at least five times using a
disposable syringe (20 ml) and a suitable irrigation
adapter.
Drain any remaining water fully.
Stage V
Dry the product thoroughly with a lint-free cloth or
medical compressed air.
15