Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine activ C FW645R Gebrauchsanweisung/Technische Beschreibung Seite 14

Handgriff, fräserführung, probeimplantate
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
Aesculap Spine
activ C handle, reamer guide, chisel guide, trial implants
Manual cleaning/disinfection
Prior to manual disinfecting, allow water to drip off
for a sufficient length of time to prevent dilution of
the disinfecting solution.
Note
3
Clean reamer guide
, chisel guide
11
with locking pipes
in closed position (
move them during cleaning.
After manual cleaning/disinfection, check visible
surfaces visually for residues.
Where necessary, repeat the cleaning process.
Manual cleaning with immersion disinfection
FW789R–FW800R
Stage
Step
I
Cleaning
II
Intermediate
rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at
least)
RT:
Room temperature
12
10
2
and handle
) and do not
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
2
(cold)
RT
1
-
(cold)
RT
15
2
(cold)
RT
1
-
(cold)
RT
-
-
Water
Chemical
quality
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QAV-free
D–W
-
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QAV-free
DM-W
-
-
-

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Aesculap spine activ c fw647r

Inhaltsverzeichnis