Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine activ C FW645R Gebrauchsanweisung/Technische Beschreibung Seite 16

Handgriff, fräserführung, probeimplantate
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
Aesculap Spine
activ C handle, reamer guide, chisel guide, trial implants
Manual cleaning with ultrasound and immersion disinfection
FW870/FW871R/FW784R
Stage
Step
I
Ultrasonic cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at
least)
RT:
Room temperature
Stage I
Clean the product in an ultrasonic cleaning bath
(frequency 35 kHz) for at least 15 minutes. Ensure
that all accessible surfaces are immersed and
acoustic shadows are avoided.
Clean the product with a suitable cleaning brush
until all discernable residues have been removed.
Use a suitable cleaning brush (TA011944) to clean
all surfaces which are not visible, e.g. in
instruments with concealed crevices, lumens or
complex geometries, for at least 1 min or until no
further residues can be removed.
Mobilize non-rigid components, such as set
screws, joints, etc. during cleaning.
14
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
2
(cold)
RT
1
-
(cold)
RT
15
2
(cold)
RT
1
-
(cold)
-
-
-
After cleaning, use a 20-ml single-use syringe to
rinse thoroughly, for at least 5 times, these difficult
to access parts of the product.
Do not use metal cleaning brushes or other
abrasives that would damage the product surfaces
and could cause corrosion.
Stage II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Drain any remaining water fully.
Water
Chemical
quality
D–W
BBraun Stabimed; phenol-
free, aldehyde-free and
QAV-free
D–W
-
D–W
BBraun Stabimed; phenol-
free, aldehyde-free and
QAV-free
DM-W
-
-
-

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Aesculap spine activ c fw647r

Inhaltsverzeichnis