Herunterladen Inhalt Inhalt Diese Seite drucken

College Park voyager Technische Anleitung Seite 4

Inhaltsverzeichnis

Werbung

ASSEMBLY AND DISASSEMBLY
EN
(FOR SOCK REPLACEMENT)
Use the Foot Horn to don and doff the foot shell. Remove the CPI sock and replace as needed.
Any further disassembly or modification of components will void the warranty.
ENDOSKELETAL MOUNTING
Use only high quality proximal endoskeletal components.
STATIC ALIGNMENT
For optimal function, balance the patient's weight evenly between the heel and toe. (Figure 2)
▪ The Voyager was designed with a 3/8" (10 mm) heel rise.
▪ The load line divides the foot at 1/3 heel lever and 2/3 toe lever.
DYNAMIC ADJUSTMENTS
DESIRED RESULT
Firmer toe response
Softer toe response
Firmer heel response
Softer heel response
*To install or remove heel wedge, refer to the Voyager Heel Wedge Kit Instructions.
TORSION ADJUSTMENTS
▪ Range of adjustment = 2 full turns
▪ Factory setting = minimum resistance
SYMPTOM
Rotation too fast
Rotation too slow
WARNINGS
WARNING
▪ Do not expose this product to corrosive materials, salt water or pH extremes.
▪ Any further disassembly or modification of components will void the warranty.
Failure to follow these technical instructions or use of this product outside the scope of its Limited Warranty
may result in injury to the patient or damage to the product.
ALIGNMENT CHANGE
Plantarflex the Voyager or move
load line posterior
D
orsiflex the Voyager or move load
line anterior
D
orsiflex the Voyager or move load
line anterior
Planta
rflex the Voyager or move
load line posterior
(FIGURE 3, TOP OF FOOT)
DESIRED RESULT
Less torsion moment/
Increase Resistance
Greater torsion moment/
Decrease Resistance
COMPONENT CHANGE
---
---
Add heel wedge
Remove heel wedge
ADJUSTMENT
Turn clockwise
Turn counterclockwise