Oscillating saw straight GA336 / Oscillating Sternum saw GA337
Issue
Cause
Insufficient cutting
Saw blade teeth worn
performance of the
Battery too weak
saw blade
Battery worn/defective
Saw blade overheated
The saw blade cannot
Coupling components or saw
be coupled
blades deformed
-
Saw blade cannot be
Locking pin not fully releasing
uncoupled
the saw blade

10. Technical Service

Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
14
Acculan 4
®
Detection
Blunt saw blade
Insufficient motor power/speed
Inadequate chip removal
Deformation
Incompatible saw blade
Button for saw blade release not
pressed completely
Remedy
Replace saw blade
Charge battery
Replace battery
Move the saw back and forth during sawing and
ensure sufficient irrigation of the saw blade
Have repaired by manufacturer
Insert a compatible Aesculap saw blade, see

Accessories/Spare parts

Fully press the saw blade release button, see
Attaching and detaching the saw blade
11. Accessories/Spare parts
Art. no.
Designation
GA675
Cover
GA676
NiMH battery long
GA678
Sterile funnel
GA679
Battery removal device
GB600
STERILIT Power Systems oil spray
GB498R
ECCOS fixation for straight machine
GB487R
ECCOS fixation for lid
GB488R
ECCOS fixation for sterile funnel
GB489R
ECCOS fixation for battery removal device
TA014546
Instructions for use of oscillating saw straight
GA336 / Oscillating Sternum saw GA337 (A4 for
ring binder)
TA014547
Instructions for use of oscillating saw straight
GA336 / Oscillating Sternum saw GA337 (flyer)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Aesculap acculan 4 ga337

Inhaltsverzeichnis