Herunterladen Inhalt Inhalt Diese Seite drucken

Endress+Hauser Proline Prosonic Flow P 500 Installationsanleitung Seite 24

Austausch sensor, sensorhalterung, installations set
Vorschau ausblenden Andere Handbücher für Proline Prosonic Flow P 500:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EA01368D
8.3
Installing sensor – small nominal diameters
DN 15 to 65 (½ to 2½")(Prosonic Flow P, W)
Requirements
• The installation distance is known →  9
• The sensor holder is pre-assembled
Material
The following material is required for mounting:
• Sensor incl. adapter cable
• Sensor cable for connecting to the transmitter
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the
sensor and pipe
Procedure:
1.
Set the distance between the sensors to the value determined for the sensor distance.
Press the movable sensor down slightly to move it.
 9
Distance between sensors as per the installation distance →  9
a
Sensor distance (back of sensor must touch the surface)
b
Sensor contact surfaces
2.
Stick the coupling pad under the sensor to the measuring pipe or coat the contact
surfaces of the sensor (b) with an even layer of coupling gel (approx.
0.5 to 1 mm (0.02 to 0.04 in)).
24
a
b
Endress+Hauser

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Prosonic flow i 400Prosonic flow w 400

Inhaltsverzeichnis