Sensor holder with U-shaped screws (Prosonic Flow P, W)
Can be used for
• Measuring devices with measuring range DN 15 to 65 (½ to 2½")
• Mounting on pipes DN 15 to 32 (½ to 1¼")
Procedure:
1.
Disconnect the sensor from the sensor holder.
2.
Position the sensor holder on the measuring pipe.
3.
Fit the U-shaped screws through the sensor holder and lightly grease the thread.
4.
Screw the nuts onto the U-shaped screws.
5.
Position the sensor holder correctly and tighten the nuts uniformly.
 2
Holder with U-shaped screws
1
Sensor holder
CAUTION
L
Risk of damaging plastic or glass pipes if the nuts on the U-shaped screws are tightened
too much!
The use of a metal half-shell (on the opposite side of the sensor) is recommended for
plastic or glass pipes.
The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to
ensure good acoustic contact.