Herunterladen Inhalt Inhalt Diese Seite drucken

WIKA GDM-RC-100-T Betriebsanleitung Seite 28

Hybrid-gasdichtewächter mit referenzkammer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Commissioning, operation
5.6.2.6 Configuration
Cross-check the as-delivered condition of the configuration with the delivery note. The
configuration ex-works may differ from the standard described here.
Register
Parameter
EN
00100
Address
00101
Baud rate
Parity
00102
00106
Serial number
00110
HW version
00111
SW version
Model description
00112
00113
TAG number (name of gas
density sensor)
00160
Gas mixture SF
00161
Gas mixture N
00162
Gas mixture CF
Gas mixture O
00163
00164
Gas mixture CO
00165
Gas mixture Novec 4710
00166
Gas mixture He
Gas mixture Ar
00167
Address
The available address space is 1 ... 247 (standard 247).
Baud rate
The different speeds are presented with register values 0 ... 8.
Baud rate
1,200
2,400
4,800
9,600
14,400
19,200
38,400
57,600
115,200
28
6
2
4
2
2
Register value
0
1
2
3
4
5 (standard)
6
7
8
Value definition
Standard
1 ... 247
247
1,200 ... 115,200
19,200
None, Even
None
2 = model GD-20-D
16 byte ASCII
0 ... 100 %
100 %
0 ... 100 %
0 %
0 ... 100 %
0 %
0 ... 100 %
0 %
0 ... 100 %
0 %
0 ... 100 %
0 %
0 ... 100 %
0 %
0 ... 100 %
0 %
WIKA operating instructions, model GDM-RC-100-T
Writable
Yes
Yes
Yes
Read only
Read only
Read only
Read only
Yes
Read only
Read only
Read only
Read only
Read only
Read only
Read only
Read only

Werbung

Kapitel

Inhaltsverzeichnis
loading