6. Faults
Faults
Contact is no longer
switching in accordance
with the specification
EN
Switching status remains
unchanged despite reach-
ing the switch point/reset
point
No pointer movement
despite change in pressure
Pointer movement, even
though depressurised
Gas density falls continu-
ously
No communication via
Modbus
or current signal
®
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the gas
density sensor must be taken out of operation immediately.
For claims, the serial and product numbers must be stated. The serial number is printed
on the dial, the product number on the product label. For claims, the atmospheric
pressure and the temperature during the measurement must be given, as well as the
data on the reference standard (model, class).
32
Causes
Electrical connection is inter-
rupted
Electrical load unsuitable for
the switch contact model
Contact contaminated
Contacts defective (e.g. fused
contact zone)
Movement blocked
Warming or cooling of the
measuring instrument (no
fault)
Leakage in the gas compart-
ment
Electrical connection not
correct
Configuration error
Ensure that pressure or signal is no longer present and protect against
accidental commissioning. Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter 8.2
"Return".
Measures
Carry out a continuity test on
the electrical connection leads
Maintain the permissible
electrical loads for the switch
contact model
Replace instrument. Before
recommissioning the new
instrument, provide a protec-
tive circuit for the contact
Replace instrument
Let the instrument settle for 2
hours at 20 °C [68 °F]
Check the mechanical
installation of the measuring
instrument
Search for leaks with leak
detector e.g. model GIR-10
Check wiring and power
supply
Query via WIKA start-up kit
WIKA operating instructions, model GDM-RC-100-T