5. Commissioning, operation
5.5 Modbus
®
The Modbus
communication protocol is based on a master/slave architecture. The
®
protocol implemented in the model GD-20 gas density sensor is Modbus
transmission via a 2-wire RS-485 interface.
The Modbus
protocol is a single-master protocol. This master controls the entire data
®
transfer and monitors any possible timeouts (no reply from the addressed instrument).
The connected instruments may only send messages after request by means of the
master.
Modbus
-RTU (RTU: Remote Terminal Unit) transmits the data in binary form, guaran-
®
teeing a good data throughput.
Detailed information on the protocol under www.Modbus.org
5.6 Modbus
start-up kit
®
The gas density sensor, with the optionally available start-up kit (order no. 14075896),
can be configured for operation at the measuring location.
A further function is an integrated data logger, showing measured data in a specific
cycle or writing it in a file.
The start-up kit consists of:
Power supply unit
Interface converter (RS-485 to USB)
USB cable type A to type B
Sensor cable with M12 x 1 connector
Adapter cable for models GDM-100-T and GDM-RC-100-T
Modbus
tool
®
5.6.1 Establish connection to the PC
Interface converter
RS-485
USB
+
-
WIKA operating instructions, model GDM-RC-100-T
Voltage supply
-RTU with serial
®
EN
23