Hand-Held Pressure Indicator
CPH6200-Ex (Atex-Version)
GB
8
As is normal for all measuring instruments, the accuracy of the device
should be tested at regular intervals (approx. once per year - see Sec-
tion 8).
1.4 Electrical connections
On the upper edge of the device:
The connector socket for attaching the CPT62I0 pressure sensors, is
located on the top edge of the CPH6200-Ex (see Section 6), as is the
socket for the WIKA interface cable (see Section 5).
Connector socket for CPT62I0-Sensorcable
1.5 Display
7
8
1. Main display: shows the current measured value of the connected
pressure sensor.
2
: indicates a low battery and other warnings.
3. Lower display: dispalys the min, max or hold function values.
4. Logg: shown if logger function is selected, flashes if cyclic logger
is running.
5. AL: flashes if alarm is present.
6. Tare: indicates that tare function is activated.
7. Measuring unit: an arrow points to the chosen measuring unit.
8. SL: indicates, that the sea level function is activated.
WIKA Operating Instructions Hand-Held Pessure Indicator
Connection for WIKA inter-
face cable (do not use in
explosive atmosphere)
1
2
3
4
5
6