4. Design and function
The battery indicator lights up
To avoid false readings, replace the batteries.
For instructions on batteries see chapter 8.1.1 "Battery replacement".
EN

4.4 WIKA-Wireless

To start wireless data transmission, you must set it to "On" under "Menu / Basic
settings / Wireless". Once this has been done, the wireless symbol blinks in the display.
As soon as the CPG1500 is connected to a computer via this wireless interface, the
symbol lights continuously.

4.5 Connections

The CPG1500 is available with all standard industrial threaded connections, as standard
G ½ B is specified.
When screwing the NPT thread of the CPG1500 to an adapter or a pressure connection,
it is necessary to use additional sealing material between the threads, e.g. PTFE tape.
The hose, lines and fittings etc. must always be approved at least for the working pressure
which corresponds to the pressure of the instrument. In addition, there must be no leaks
during calibration process - if necessary, seal using PTFE tape.

4.6 Protective rubber cap for case

Optionally, the CPG1500 can be fitted with an impact-resistant protective rubber cap.
WIKA operating instructions precision digital pressure gauge CPG1500
17