Herunterladen Inhalt Inhalt Diese Seite drucken

WIKA CPH6200-Ex Betriebsanleitung Seite 12

Explosionsgeschütztes hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH6200-Ex:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Hand-Held Pressure Indicator
CPH6200-Ex (Atex-Version)
Menu
Param. Setting
GB
„Menu"
SEt
Set Configuration: Generic settings
ConF
Unit
SL
Alti
rAtE
t.AVG
P.oFF
Adr.
SEt
Set Alarm: Alarm function configuration
AL.
AL.
AL.Lo
AL.Hi
SEt
Set Logger: Logger function configuration
LoGG
Func
CYCL
Lo.Po
SEt
Set Clock: Setting of real time clock
CLOC
CLOC
dAtE
YEAr
12
Meaning
or
mbar, bar, ...
Unit: Unit of display
oFF / on
Sea-level correction: on or off
-200 ... +9999
Altitude: Input of altitude above sea-level [m] (only if on)
Rate: Measuring rate (see chapter 2.3)
Slo
Slow: Measuring rate (4 Hz filtered, low power consumption)
FASt
Fast: Measuring rate, filtered (> 100 Hz)
P.dEt
Peak detection: fast measuring rate, unfildered (> 100 Hz)
1 ... 120
Averaging interval in seconds, used by the averaging function off
oFF
Averaging function deactivated
1 ... 120
Auto Power-Off delay in minutes. Device will automatically switch itself off if,
during this interval, no key is pressed, or there is no serial communication.
oFF
Auto Power-Off function inactive (continuous operation)
01, 11 ... 91
Communicationsaddress of interface
on
Alarm function on
no. So
Alarm function on
oFF
No alarm function
Min-Range ...
Min alarm limit
(not when AL.oFF, Min-Range is the lower display range of the connected
... AL.Hi
sensor)
AL.Lo ...
Max alarm limit
(not when AL.oFF, Max-Range is the upper display range of the connected
... Max-Range
sensor)
CYCL
Cyclic: logger function 'cyclic logger'
Stor
Store: logger function 'discrete value logger'
oFF
No logger function
1 ... 3600
Cycle time for cyclic logger [seconds]
on / oFF
Low-power logger with very low power consumption (only if cyclic logger with
slow measuring rate)
HH:MM
Clock: set time hours:minutes
TT.MM
Date: set time day.month
YYYY
Year: set time year
* If there is existing data in the logger memory, the menus/parameters
cannot be accessed! If these need to be adjusted, the logger
memory must first be cleared! (see Section 4)
WIKA Operating Instructions Hand-Held Pessure Indicator
*
*
*
*
*
*
*
*
*
*
*
*
*
*

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph62i0

Inhaltsverzeichnis