Operation of min/max-value memory in conjunction with the averaging
function:
If averaging is activated and 'slow' measuring rate is selected
(rAtE-Slo), the min-/max-memory value relates to the average
display value.
If averaging is activated and 'fast' measuring rate is selected
(rAtE-FASt or -P.dEt), the min-/max-memory value relates to the
internal measured values (pressure spikes can be detected).
2.5 (P.oFF) Power off function
If no key is pressed, and no serial communication occurs during the
interval set in configuration (P.off), the device will automatically switch
itself off to save battery power.
If P.oFF = oFF then the automatic power off is deactivated.
2.6. (Adr.) Serial communications address
This parameter has been included in preparation for future operation
with multi-point monitoring systems.
2.7 (AL.) Alarm
There are three possible settings: Alarm off (AL.oFF), on (AL.on), or
(AL.no.So).
The following conditions will prompt an alarm to be displayed, if the
alarm function is activated (AL.on or AL.no.So):
Value is below lower alarm limit (AL.Lo) or above upper alarm
limit (AL.Hi).
Sensor error (Sens-Erro)
Low battery (bAt)
Fe 7: System error
When an alarm occurs, the alarm arrow flashes, and, during serial
communication, the 'PRIO'-flag is set within the transmitted message.
activated channels.