Hand-Held Pressure Indicator
CPH6200-Ex (Atex-Version)
GB
20

7. Fault and system messages

Display
Meaning
Low battery power, device will only
work for a short while longer
Battery flat
SEnS
No sensor connected
Erro
Connected sensor or device faulty If second sensor available, check
- - - -
Logger data is being read by the
serial interface
- - - -
No display
Battery flat
or strange
symbols,
System error
device is
not res-
ponding to
Instrument faulty
input
Err.1
Measured value above allowable
range
Sensor faulty
Err.2
Measured value below allowable
range
Sensor defective
Err.4
Value is to low to be displayed,
tare is set
Err.7
System error
Err.9
Connected sensor or device faulty If second sensor available, check
Measured value is far above or
below allowable range
WIKA Operating Instructions Hand-Held Pessure Indicator
Remedy
Replace battery
Replace battery
Switch off device and connect
sensor
if device is ok. Return defective
device/sensor to manufacturer
for repair
When the data transfer is com-
pleted the device will automati-
cally return to normal measuring
display, no remedy necessary
Replace battery
Disconnect battery, wait a while,
re-connect
Return to manufacturer for repair
Check: Is the pressure outside
the sensor's range? -> measured
value to high!
Return to manufacturer for repair
Check: Is the pressure outside
the sensor's range? -> measured
value to low!
Return to manufacturer for repair
Check: is value below '-2000'
(check Tare)?
Return to manufacturer for repair
if device is ok. Return defective
device/sensor to manufacturer
for repair
Check: Is the pressure outside the
sensor range?

Werbung