Certificate: 70009819
Project: 80065116
Linear
Maximum permissible external values for common effective reactance's (C
Single reactance's to table A.2 and figure A.4 or A.6 of 60079-11
L
o
17.0
Or
HART-current output (Sink Mode) (Terminal 17/19+, 18/20-)
Type of protection Intrinsic Safety Ex ia IIB/IIC/IIIC Maximum Values:
U
= 30.0 V;
i
I
= 152.0 mA
i
P
= 1.14 W
i
Linear
*Note: the HART current output (source mode or sink mode) of the current output module are safely electrically
isolated from earth and all other circuits.
DQD 507 Rev. 2019-04-30
SENSseries LB 480
BERTHOLD TECHNOLOGIES GmbH & Co. KG
L
(mH)
o
0.44
0.8
1.6
13.0
IIB
(mH)
C
(µF)
o
0.820
L
= 20.0 µH
i
C
= 3.0 nF
i
IIB (IIIC)
C
(µF)
o
0.52
0.45
0.38
0.37
IIC
L
(mH)
o
4.0
© 2018 CSA Group. All rights reserved.
Master Contract: 215040
Date Issued: 2021-09-27
is not considered)
i
IIC
C
(µF)
o
0.084
0.066
0.049
-
C
(µF)
o
0.107
Page 8
1 – 81

Werbung