Fehlermeldung
101
HW module missing or not tested
103
RAM Error
104
Device Error
108
Safety parameter invalid
200
Data flow
201
Error by analog input calibration
202
Clock signal deviation
300
Data flow
301
Error by ADC calibration
302
Error by DAC calibration
303
Supply 5.0V
304
Reference 2.0V
305
Reference 2.5V
306
ERROR GND CPU ST9
307
No impulses in measuring channnel
308
No impulses in control channel
309
No impulses in auxiliary channel
310
Impulse difference measuring channel
311
Impulse difference control channel
312
Impulse difference auxiliary channel
313
Instable pulse rate
314
Threshold of measurement channel 1
315
Threshold of measurement channel 2
316
Threshold of control channel 1
317
Threshold of control channel 2
318
Threshold of auxiliary channel 1
319
Threshold of auxiliary channel 2
330
Detector malfunction
400
Data flow
401
Supply 11V
402
Supply 5V
403
Supply 5VM
404
Supply 3,3V
500
Data flow
502
Digital input malfunction
503
Digital output malfunction
600
Data flow
700
Data flow
701
Impulse difference
704
ERROR GND CPU ST6
900
Data Flow
SENSseries LB 480 Füllstandsmessung
BERTHOLD TECHNOLOGIES GmbH & Co. KG
8.2.2
Fehlerbehandlung
Fehlergrund
Hardwarefehler
8 Fehlerbehandlung
Fehlerbehandlung
Der Fehler lässt sich nur durch Tausch
des Detektors oder der Detektorelektro-
nik beheben.