Installation

DANGER
Before installing or removing the P3 pressure transducer, you must check that
the line is not pressurized.
The pressure transducer can be screwed in wherever required. If the trans­
ducer is used to measure dynamic pressure characteristics in liquids, it should
be installed with the pressure connection pointing upward, so that it is not pos­
sible for an air cushion to build up in the measurement tube.
The connecting pin with the M 12 x 1.5 and M 20 x 1.5 threaded connectors for
the P3/3000 bar type, corresponds to DIN 16288.
Notice
When tightening, the wrench (27 a.f.) must only be put to the flat of the mount­
ing flange and not to the housing or to the cable entry. The permissible tighten­
ing torque is 30 NVm.
Pay particular attention to the sealing on the thread of the connecting pin. The
pressure medium and the respective conditions of installation for the individual
case will determine which type of sealing is suitable. Some of the sealing
options are shown below.