5.1 Simulation

A simulation allows the functions of the electronic insert to be simulated and tested.
Field
Parameter
V9H6
Simulation
disabled - enabled
V9H7
Simulation value
When simulation field V9H6 is enabled, the last process variable value is
automatically loaded in V9H7 field and applied to V0H0 (hold condition).
The V9H7 value can be manually modified to simulate the process variable.
Warning E613 appears in field V9H0 during simulation.

5.2 Fault codes

With the PV value a diagnostic code is sent on PROFIBUS-PA fieldbus, displayed in
matrix field V9H8. The displayed value may be a single fault code or a sum of single
codes.
E.g. code 5 is the sum of codes: 4 and 1.
The significance of each single code is shown in the table below.
Code
Significance
0
No fault
1
Input failure
2
Input under-range
4
Input over-range
64
Cold junction error
128
Hardware error
5.3 Reset to factory setting
The TMD 834 allows two different resets:
• Code 7864:
Reset of all parameter to factory settings
• Code 333:
Reset of all parameter to factory settings, with the
exception of bus address.
These codes may be entered in field V9H5.
On reset to factory parameters (Code 333), values in the table below are assumed.
The values in the grey fields are retained to the customer setting.
H0
H1
V0
V1
0/400 Ω Pt100
°C
V2
- 200 °C +850 °C
V9
Set input simulation on or off. See also next field V9H7.
The input simulation value in engineering units must be entered
here if simulation is enabled in field V9H6.