NOTES ON THE ARM SYSTEMS

18.1

Cleaning and disinfection

DANGER
Electric shocks are life-threatening.
The devices can conduct electricity and must be
handled with care during cleaning and disinfection.
• If there is a power plug remove it from the socket.
• Do not use spraying methods for cleaning and/or
disinfection.
• Do not spray liquid into the openings of electrical
power sockets, gas sockets or devices, or allow liquid
to enter these.
Recommended cleaning agents
Use mild soap solutions or standard detergents as cleaning
agents. Wipe down the surfaces of the equipment with a
dampened cloth. If necessary, use small quantities of mild
soap solution as a cleaning agent. Then wipe the outer
surfaces with a soft, clean cloth until they are completely
dry.
Disinfection procedure
Disinfection by wiping is intended to be the standard
disinfection procedure for the support arm system. Hygiene
guidelines and corresponding safety measures must be
determined by the operator for the disinfection procedures
that are applied.
• After the contamination by any potentially infectious
material (such as blood, secretions or excrement), the
areas must be disinfected immediately in a targeted way.
• Comply with application concentrations.
• Wipe the area to be disinfected rather than spraying it.
• Do not use areas that have been wiped until the
disinfectant is dry.
18.2

Testing

DANGER
Electric shocks are life-threatening.
During all testing work, disconnect the device from the
power supply/remove the plug from the power socket
and secure the device so that it cannot be switched on
again.
Annually:
• Paint damage
• Cracks in plastic parts
• Deformation of the support system
• Loosening of parts
• Check the fastening screw on the underside of the stand
base, tighten it if necessary.
Every 2 years:
The securing element of the cantilever arm must be
checked and greased by a qualified professional
18.3

Demounting for servicing

DANGER
Injury hazard
The cantilever arm is under a high spring load. If the
luminaire head is not moved to the upper cantilever arm
40
position, the cantilever arm will spring up, possibly
causing serious injury. Only remove the luminaire head
if the cantilever arm is in the upper position.
1. Disconnect the device from the power supply/remove
the plug from the mains and ensure it cannot be
switched on again.
2. Place cantilever arm in the upper position.
3. Unscrew the brake screw (1) so that the sleeve (2) can
be turned.
4. Turn the sleeve (2) through 180 degrees until the
securing segment (3) appears in the slot (4).
5. Hold the luminaire head (5) firmly and remove the
securing segment (3) using a narrow flat-bladed
screwdriver.
Remove luminaire head (5) from the connection.
6. Put protective cap in place.
1 Brake screw
2 Sleeve
3 Securing segment
4 Slot
5 Luminaire head terminal