TROUBLESHOOTING

Fault
The luminaire does not
light
The luminaire does not
light
The luminaire does not
light
The luminaire does not
light
17

TECHNICAL DATA

Electrical:
Nominal input voltage
Frequency range
Power consumption
Input current
Power factor
Integrated electronic transformer
Approval
Illumination*:
Central illuminance at 1.0 m (3.28 feet)
Light field diameter d10 at 1.0 m (3.28 feet)
Light field diameter d50 at 1.0 m (3.28 feet)
Colour temperature
Colour rendering index Ra
Colour rendering index R9
Total irradiance Ee at maximum intensity
Ratio irradiance Ee to illuminance Ec
Photobiological safety (EN/IEC 60601-2-57)
Residual light intensity when shading using one shade
Residual light intensity when shading using two shades
Light intensity in tube
Residual light intensity in tube with one shade
Residual light intensity in tube with two shades
Depth of illumination L1 + L2
*Illumination data -10% / +20% tolerance
Environmental conditions for transport, storage and operation:
Ambient temperature (storage and transport)
Ambient temperature (operation)
Relative humidity (non-condensing)
Weight:
38
Possible cause
Contact problem
Lamp defective
No mains voltage
1 A fuses defective
(Polaris 50 Mobile only)
Troubleshooting
Switch on again
Contact manufacturer
service
Check mains voltage, check
all connections
Replace fuses
(2 pcs. are fitted)
User profiles
All
By manufacturer service only
Electrician
Electrician
100 - 240 VAC
50 / 60 Hz
30 W (42-52 VA)
0.42A
0.69
24 V DC output
ETL / cETL
60,000 lux
Ø = 16 cm (6.3")
Ø = 8.5 cm (3.3")
4500 K
95
90
2
200 W/m
2
3.5 mW/m
/lx
Risk Group 1
(low risk)
< 1%
56%
96%
< 1%
55%
120 cm (47.2")
-20 °C to +70 °C (-4 °F to +158
°F)
10 °C to +35 °C (+50 °F to +95 °F)
max. 75%