AT235 Instructions for Use - US

4.3 Cleaning the Probe Tip

Step 2: Thread the stiff end of the cleaning brush into one of the tubes from inside. Pull the cleaning
floss completely through the probe tip tube. Clean each of the three tubes. Discard floss after use.
Notice:
Only insert the cleaning brush from inside out, this will ensure that the dirt is pushed out of
the probe instead of into the probe and as well protect the gasket from
being damaged. Never clean inside the probe holes.
Diagnostic probe
Step 1: Unscrew the probe cap and remove the probe tip.
Step 3: Replace the cleaned tip.
Step 4: Assemble the probe again.
Clinical probe