Herunterladen Inhalt Inhalt Diese Seite drucken

Application Setup - OHAUS Defender 5000-Serie Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Defender 5000-Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Defender 5000 Indicators

4.3.3 Application Setup

The Application can be customized for various user preferences.
Press the Softkey corresponding to the icon
to enter Configuration Setup
The Configuration Menu is now displayed.
Select the list item and press the Softkey
corresponding to the icon
setting as desired.
To return to the Application home screen, press
the button corresponding to the icon
The Check Configurations are defined below (defaults in Bold).
Option (Bold is default)
Configure Item
Check Mode
Weighting/Counting
Auto Tare
Off/On/Accept
Accumulate
Off/Manual/Automatic
Audible Signal
Off/Under and
Over/Accept/Under/Over
Auto Opt.
Off/On
Opt. Beep
Off/On
APW Auto Save
Off/On
.
,
to change the
.
Description
Weighing: Check weighing mode.
Counting: Check counting mode.
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Accept: If the object weight is in the range of the Over and
Under Limit you set, auto tare will be performed.
Off: The Softkey "∑" does not display.
Manual: The Softkey "∑" displays. User can press the key
to accumulate the stable weight.
Automatic: The Softkey "∑" displays. The weight will be
accumulated automatically.
Note:
The load to be accumulated has to be >= 10d and the next
accumulation can only start once the pan has been cleared.
When LFT is ON(
approve mod is OIML)
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross
0 or net 0.
Off: No beep.
Under and Over/Accept/Under/Over: Beep when
reaching the selected check point.
Off: Auto Opt. is off.
On: The APW will be optimized automatically when doing
count weighing.
Off: Opt. Beep is off.
On: When the APW is optimized, the beeper will beep
once.
Off: APW Auto Save is off.
On: If the APW is get from sampling, and a counting
library is selected, the new APW will save to library after
optimized.
no such limit when LFT is OFF or the
,
EN-43

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Diese Anleitung auch für:

Defender td52xwDefender td52p

Inhaltsverzeichnis