5.2.6

Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the scale, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.
To cancel at any time, press Cancel.
5.2.7

Calibration Test

Use Calibration Test to compare a known calibration weight against the stored span calibration data.
With no load on the scale, press Cal Test to begin the process.
The balance captures the zero point, then prompts for the span value.
The display shows status, followed by the difference between the current calibration weight and the
stored calibration data.

5.3 Balance Setup

Enter this sub-menu to customize Balance functionality.
5.3.1

Balance Set-up sub-menu

Factory default settings are shown below in bold.
Find Quality Products Online at:

Language

User Settings
Auto Tare
Graduations
5.3.2
Language
Set the language displayed for menus and displayed
messages.
English
German
French
Spanish
Italian
Polish
Turkish
Czech
Hungarian
GlobalTestSupply
www.
ADVENTURER BALANCES
Filter Level
Auto Zero
Tracking
Date & Time
Approved Mode
.com
sales@GlobalTestSupply.com

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis