EN-42
The Check Configurations are defined below (defaults in Bold).
Configure Item
Option(Bold is default)
Check Mode
Check Weighting/ Check
Counting
Auto Tare
Off/On/Accept
Accumulate
Off/Manual/Automatic
Audible Signal
Off/Under and
Over/Accept/Under/Over

4.3.2 Check Counting

Press the configuration button
Check Mode to Check Counting. Place objects on
the pan. The Under/Accept/Over status is shown
in the progress bar area while the actual number
of pieces is shown on the main Display Line.
Set the Average Piece Weight (APW)
Note: It is recommended that the APW is larger than 1d. If APW is between 0.05d and 1d, a warning
message will be displayed and the information line will show 'APW is low!'. If APW is less than
0.05d, an error message will appear and the APW value cannot be stored.
There are three methods to set the APW, see section 4.2.2 for instructions.
Defining Over/Under Limits
Press the button Edit Limit to define the limit for
counting.
Note: See section 4.3.1 for information on how to
set the Over/Under limits.
Description
Check weighing mode
Check counting mode
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Accept: If the object weight is in the range of the Over and
Under Limit you set, auto tare will be performed.
Off: The icon "∑" does not display.
Manual: The icon "∑" displays. User can press the key to
accumulate the stable weight.
Automatic: The icon "∑" displays. The weight will be
accumulated automatically.
Note:
The load to be accumulated has to be >= 10d and the
next accumulation can only start once the pan has been
cleared.
When LFT is ON(
the approve mod is OIML)
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross
0 or net 0.
Off: No beep.
Under and Over/Accept/Under/Over: Beep when reaching
the selected check point.
and select
Defender 5000 Indicators
no such limit when LFT is OFF or
,